RAW MATERIAL

Melt flow rate (MFR)

ISO 1133-1:2022 · procedure A (mass measurement)

STANDARDISED

A material fingerprint and an indicator of consistency between batches.

Purpose and principle

Measure the mass of molten polymer extruded under a specified temperature and load. The MFR characterises flow behaviour under the specified conditions.

Conditions / parameters

MaterialPP-R
Temperature / load230 °C / 2.16 kg
EquipmentExtrusion plastometer and precision balance
Measurement4 cuts · 120 s period

Test setup

Indice de fluidité MFR : schéma du plastomètre d’extrusion.
Melt flow rate (MFR) · extrusion plastometer test setup.

Test procedure

  1. Prepare the sample, clean the barrel and stabilise the heating.
  2. Load the polymer and observe the preheating time.
  3. Collect four successive extrudate segments and weigh them.
  4. Calculate each MFR, examine the spread and report the mean.

Expression / calculation

MFR = 600 × m / t   [g/10 min]   ·   m (g), t (s)

Technical relevance